DescriptionThe TE 76 is a two-station, vertical axis, long stroke reciprocating tribometer, whi
WhatsApp: +86 13816217984
E-postadress:info@qinsun-lab.com
The STD-TE 76 is a two-station, vertical axis, long stroke reciprocating tribometer, which combines the reciprocating drive mechanism and fixed specimen mounting from the TE 77 High Frequency Friction Machine with a UHV test chamber.
Contact Configurations: | Point Contact | |
Line Contact | ||
Area Contact | ||
Load Range: | 5 to 500 N | |
Laboratory Chiller: | -35°C to 150°C | |
Temperature Sensor: | PT-100 | |
Frequency Range: | 2 to 25 Hz | |
Stroke Range: | See following tables | |
Friction Transducer: | Piezo-Electric Type | |
Interface: | Serial Link Interface Module | |
Software: | COMPEND 2000 | |
Motor: | 1.1 kW a.c. vector motor, 2048 ppr encoder | |
Stroke Range | ||
Continuously Variable Cam – 0 to 12.5 mm | ||
Angle – degrees: | Minimum – mm | Maximum – mm |
0 | 0 | 2 |
18 | 1.04 | 3.04 |
36 | 2.65 | 4.65 |
54 | 4.25 | 6.25 |
72 | 5.75 | 7.75 |
90 | 7.09 | 9.09 |
108 | 8.24 | 10.24 |
126 | 9.17 | 11.17 |
144 | 9.85 | 11.85 |
162 | 10.26 | 12.26 |
180 | 10.4 | 12.4 |
Step Variable 0 to 12.5 mm: | ||
Angle – degrees: | Nominal Stroke – mm | |
0 | 0 | |
18 | 1.94 | |
36 | 3.83 | |
54 | 5.63 | |
72 | 7.29 | |
90 | 8.77 | |
108 | 10.03 | |
126 | 11.05 | |
144 | 11.79 | |
162 | 12.25 | |
180 | 12.5 | |
Step Variable 12.5 to 25 mm: | ||
Angle – degrees: | Nominal Stroke – mm | |
0 | 12.5 | |
18 | 13.05 | |
36 | 14.26 | |
54 | 15.97 | |
72 | 17.89 | |
90 | 19.8 | |
108 | 21.54 | |
126 | 23 | |
144 | 24.09 | |
162 | 24.77 | |
180 | 25 | |
Test Chamber | ||
Pressure: | Atmospheric to 100 Picobar (10-5 Pa) | |
Environment: | Air, inert gas or vacuum | |
Material: | 304 stainless steel | |
Surface Treatment: | Glass bead blasted | |
Viewports: | Fused Silica | |
Electrical Feedthrough: | 1 kVolts , 15 Ampere | |
Coaxial Feedthrough: | BNC through Flange | |
Thermocouple: | Type K through Flange | |
Vacuum Transducer: | Cold Cathode and Pirani | |
Laboratory Chiller | ||
Fluid Temperature: | -35°C to +150°C. | |
Automatically Controlled Parameters | Frequency | |
Belastning | ||
Temperature | ||
Chamber Pressure | ||
Provningstid | ||
Manually Controlled Parameter | Stroke | |
Measured Parameters | Friction | |
Belastning | ||
Temperature | ||
Chamber Pressure | ||
Frequency | ||
Friction Coefficient | ||
. | ||
Services | ||
Electricity: | 220/240 V, single phase, 50/60 Hz, with neutral and earth, 4.5 kW | |
Clean, dry air: | 4 cfm at 8 bar (120 psi) |
Ansvarsfriskrivning: Ovanstående innehåll är endast för referens och kommunikation mellan branschinsider, och garanterar inte dess riktighet eller fullständighet. Enligt gällande lagar och förordningar och förordningar på denna webbplats ska enheter eller individer som köper relaterade artiklar erhålla giltiga kvalifikationer och kvalifikationsvillkor.