Protective clothing heat conduction tester - Qinsun
Tillverkare av professionella provningsinstrument
Nationell hotline för samråd:+86 021-64208466

Protective clothing heat conduction tester

Overview1.1 Applicable for testing thermal protection performance of single-layer or multi-layer mat

WhatsApp : +86 13816217984

Email : info@qinsun-lab.com

Innehåll

Översikt

1.1 Applicable for testing thermal protection performance of single-layer or multi-layer materials used in protective clothing exposed to convective and radiant heat hazards. Not suitable for non-flammable materials or those prone to melting/dripping at high temperatures.
1.2 Design principle:
  a. Horizontal placement of flame-retardant samples under combined convective/radiant heat sources
  b. Exposure time recorded when transmitted heat equals second-degree burn threshold
  c. Direct contact simulation (0mm gap) vs spaced contact (6.5mm gap)
  d. Thermal protection value (TPP) calculation: Intersection of heat flux-time curve with Stoll curve
  e. Stoll curve: Standard curve relating burn time to energy exposure – values above curve indicate risk of second-degree burns

Compliance Standards

GB 8965.1-2009 Appendix A, GA 10-2014 Appendix A, NFPA 1971-2017, ISO 12749, GB/T 38302-2019

Product Advantages

1. Computerized control system with LabVIEW software integration
  a. User-friendly interface with automated calculations
  b. Real-time data acquisition (20 samples/sec) and Stoll curve intersection detection
  c. Intelligent temperature compensation algorithm
2. Automated thermal shield system:
  a. ≤0.5s response time for heat shield retraction
  b. Water-cooled shielding mechanism
3. Precision ignition system:
  a. High-voltage spark generator with safety interlock
  b. 38mm diameter gas burner with adjustable air/fuel mixture
4. TPP automatic calculation:
  a. Integration of exposure time and total heat flux
  b. Digital display of calculated TPP value

Tekniska specifikationer

SpecifikationDetails
4.1 Heat sources4.1.1 Radiant: 9×T-150 quartz lamps (10–50 kW/cm² @ 127mm)
4.1.2 Convective: Twin gas burners (45° angle, 10–50 kW/cm² each)
4.1.3 Total heat flux: 83±2 kW/cm²
4.2 SensorsCopper calorimeter probe (Φ25mm) with ±0.1℃ resolution
Stoll curve intersection detection system
4.3 Data acquisition20 Hz sampling rate prior to 250℃
Automatic TPP calculation via software integration
4.4 Safety featuresHigh-voltage interlock system
Water-cooled shield mechanism
Emergency shutdown protocol
4.5 DimensionsMain chamber: 800×500×550mm
Instrument footprint: 1200×800×1000mm
4.6 Power requirementsAC 220V±10% 50Hz, 3kW maximum

image.png

Ansvarsfriskrivning: Ovanstående innehåll är endast för referens och kommunikation mellan branschinsider, och garanterar inte dess riktighet eller fullständighet. Enligt gällande lagar och förordningar och förordningar på denna webbplats ska enheter eller individer som köper relaterade artiklar erhålla giltiga kvalifikationer och kvalifikationsvillkor.


Nätmeddelande

Kontaktuppgifter

Företagstelefon

+86-21-6420 0566

Arbetstider

måndag till fredag

Mobiltelefon:

13816217984

E-post:

info@qinsun-lab.com

QR-kod