**Scope of Application**Determines the liquid loss performance of nonwoven fabrics. Place a standard
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
**Scope of Application**
Determines the liquid loss performance of nonwoven fabrics. Place a standard absorbent medium beneath the test specimen, position the combined sample on an inclined platform, and measure the liquid penetration through the nonwoven fabric into the absorbent medium when a specific volume of artificial urine flows downward. Calculate the liquid loss performance by weighing the absorbent medium before and after testing.
**Compliance Standards**
Edana 152.0-99; ISO 9073-11
**Technical Specifications**
1. Test bench features two black reference lines:
- Low line: Located 3±0.2 mm above the bench end (absorbent medium position)
- High line: Approximately 25 mm below specimen top (aligns with drainage tube center)
2. Bench inclination: 25°
3. Clamp assembly:
- Adjustable device for fixing specimens symmetrically at (140±0.2) mm from reference line
4. Central positioning mechanism
5. Absorbent pad support frame at specimen lower edge
6. Glass tube specifications:
- Inner diameter: 5 mm
7. Ring-shaped base plate
8. Drip feeding system:
- Delivers (25±0.5) g test liquid continuously through glass tube in (4±0.1) seconds
Ansvarsfriskrivning: Ovanstående innehåll är endast för referens och kommunikation mellan branschinsider, och garanterar inte dess riktighet eller fullständighet. Enligt gällande lagar och förordningar och förordningar på denna webbplats ska enheter eller individer som köper relaterade artiklar erhålla giltiga kvalifikationer och kvalifikationsvillkor.