The anchoring force test of the guide needle can ensure the anchoring effect of the mammary positioning filament within the tissue and prevent the filament from shifting or falling out before surgery. This is essential for stable positioning filaments to accurately locate and remove the tumor during surgery.
Test standard
YY/T0880-2013 "Disposable Breast Positioning wire and Guide Needle"
Technical parameter
1. Tensile force test range: 0——20N, error is not more than +1%, resolution 0.01N
2. Test time: 1-999S, any choice, the error is not more than +0.5S
3. Test speed: 10mm/min, 50mm/min, 100m/min
4. Distance adjustment: 10mm——50mm adjustable
5. Sensor: high-precision force value sensor (with hardware and software protection function)
6. Fixture: high quality 304 stainless steel
7. Force measuring mechanism: built-in
8. Printer: on-board micro needle printer
9. Data storage: USB disk can be imported into the computer storage
10. Machine power: 70W
11. Power supply: AC90V-240V/50Hz
Ansvarsfriskrivning: Ovanstående innehåll är endast för referens och kommunikation mellan branschinsider, och garanterar inte dess riktighet eller fullständighet. Enligt gällande lagar och förordningar och förordningar på denna webbplats ska enheter eller individer som köper relaterade artiklar erhålla giltiga kvalifikationer och kvalifikationsvillkor.