Medical mask bacterial filtration efficiency (BFE) testerIt is composed of aerosol generation system
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
Tester för bakteriefiltrering (BFE) för medicinsk mask
It is composed of aerosol generation system, aerosol chamber and aerosol transmission device, negative pressure cabinet and 28.3l /min sampler.The whole tester is controlled by the console.The control console adopts microcomputer control system to coordinate and control the work of aerosol generation system, transmission system, negative pressure cabinet and sampling system, and display the working status in real time. The whole measurement work is completed automatically.
Mask Bacterial Filtration Efficiency Tester Features
Key Specification | Specification Range | Resolution | Max. permissible errors |
A Route Sampling Flow | 28.3L/min | 0.1L/min | Within ± 2.0% |
B Route Sampling Flow | 28.3L/min | 0.1L/min | Within ±2.0% |
Spray Flow | (8~10)L/min | 0.1L/min | Within ±5.0% |
Peristaltic Pump Flow | (0.006~3.0)mL/min | 0.001ml/min | Within ±2.0% |
Front Pressure of A Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.0% |
Front Pressure of B Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.0% |
Front Pressure of Spray Flowmeter | (0~300)kPa | 0.1kPa | Within ±2.0% |
Negative Pressure of the Chamber | (-90~-120)Pa | 0.1Pa | Within ±1.0% |
Working Temperature | 0-50℃ | ||
Negative Pressure of the Chamber | >120Pa | ||
Data Save Ability | >500000 sets(Scalable capacity) |
Ansvarsfriskrivning: Ovanstående innehåll är endast för referens och kommunikation mellan branschinsider, och garanterar inte dess riktighet eller fullständighet. Enligt gällande lagar och förordningar och förordningar på denna webbplats ska enheter eller individer som köper relaterade artiklar erhålla giltiga kvalifikationer och kvalifikationsvillkor.