Test PrincipleUnder low temperature conditions (e.g., liquid nitrogen bath, 77.3K), a certain a
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
Under low temperature conditions (e.g., liquid nitrogen bath, 77.3K), a certain amount of adsorbate gas (e.g., N₂) is introduced into the sample tube. The adsorption partial pressure is directly measured by controlling the equilibrium pressure in the sample tube, and the adsorption amount at this partial pressure point is obtained through the gas equation of state (PV=nRT). By gradually adding adsorbate gas to increase the adsorption equilibrium pressure, the adsorption isotherm is obtained. By gradually removing adsorbate gas to decrease the adsorption equilibrium pressure, the desorption isotherm is obtained. Compared with the dynamic method, no carrier gas (He) is required, and there is no need to repeatedly raise and lower the liquid nitrogen cup. Since the sample to be tested is in a fixed-volume sample tube, the adsorbate does not flow relative to the dynamic chromatographic method, hence the name static volumetric method.
◆ ProvningsfunktionerUtvärdering av specifik yta, porstruktur och andra gasadsorptionsegenskaper
◆ Typer av adsorbaterProvning av olika icke-frätande gaser
◆ ProvningsområdeSpecifik yta över 0,0005 m²/g, pordiameter 0,35–500 nm
◆ Antal analysstationer: 1/2 analysis station;
◆ Antal avgasningsstationer: 2 pretreatment stations;
◆ OberoendeProvningssystemet och avgasningssystemet är oberoende av varandra och möjliggör samtidig provprovning och avgasning.
◆ ProvningsnoggrannhetSpecifikt ytområde ≤±1,0%, mest sannolika repeterbarhetsavvikelse för pordiametern ≤0,15 nm, verklig densitet ≤±0,04%, yttre ytarea ≤±1,5
◆ Oberoende P0-stationPatenterad oberoende spiral mättat ångtryck (P0) teststation för att säkerställa hög noggrannhet vid partiell tryckprovning
◆ Temperaturprogrammerad avgasning: Software-controlled temperature programming, room temperature-400°C, accuracy better than 0.1°C; (customizable for special requirements)
◆ Ultimat vakuum:
PS series: Original imported two-stage mechanical vacuum pump, vacuum degree 4*10-2 Pa,
PM series: 1 original imported turbomolecular pump + 2 original imported mechanical pumps, P/P0 as low as 10-8
◆ Tryckmätning:
PS-serie: 1 original importerad trycksensor per teststation, område 0-1000 torr
PM-serien: Flera trycksensorer per teststation, intervall 0-1 torr och 0-1000 torr (0-0,1 torr tillval);
◆ Öppna datagränssnittProgramvaran kan integreras med LIMS för att ladda upp testresultat till LIMS.
Ansvarsfriskrivning: Ovanstående innehåll är endast för referens och kommunikation mellan branschinsider, och garanterar inte dess riktighet eller fullständighet. Enligt gällande lagar och förordningar och förordningar på denna webbplats ska enheter eller individer som köper relaterade artiklar erhålla giltiga kvalifikationer och kvalifikationsvillkor.